GF Signet 3-2716-WT 3-27XX-XX Paddlewheel pH/ORP Electrode & Sensors
3-27XX-XX Paddlewheel pH/ORP Electrode & Sensors for analytical process measurement
GF Signet 3-2716-WT is the 3-27XX-XX Paddlewheel pH/ORP Electrode & Sensors for analytical process measurement.
GF Signet 3-2716-WT 3-27XX-XX Paddlewheel pH/ORP Electrode & Sensors
GF Signet 3-2716-WT is the 3-27XX-XX Paddlewheel pH/ORP Electrode & Sensors for analytical process measurement.
Specifications and Features
2714-2717 pH/ORP Electrodes
Signet 2714-2717 pH and ORP Electrodes provide unsurpassed simplicity, reliability and accuracy for a wide variety of industrial applications. Rugged CPVC construction, large reference volume and intelligent positioning of internal elements combine to extend the service life of these dependable and highly responsive sensors. Flat versions allow sediment and particles to sweep past the measurement surface, minimizing risks of abrasion, breakage and coating.
The unique twist-lock design enables sensor connections in one easy motion. This includes the integral temperature sensor in pH electrodes, and the I.D. circuit in ORP electrodes. These contacts are used for automatic sensor recognition by Signet pH/ORP Instrumentation, adding convenience and versatility to our systems.
Fields of Application:
- Water and Wastewater Treatment
- Neutralization Systems
- Cooling Towers
- Effluent Monitoring
- Sanitization Systems
- Commercial Pools and Spas
- Aquatic Animal Life Support Systems
- Process Control
- Boiler Protection
Advantages:
- Durable CPVC or glass body with Twist-Lock connector
- Flat or bulb surfaced electrodes available
- Large reference volume and solid polymer electrolyte
- Integrated temperature sensor (pH)
- Designed for maximum protection from process contamination
- DI option (pH) for pure water use (
- HF option (pH) extends electrode life in applications containing trace amounts of HF (
Product Details
- Manufacturer: GF Signet
- Part number: 3-2716-WT
- Product type: pH/ORP Electrode
Review range, materials and electronics compatibility.
| Operating Range 2714, 2716, 2716-DI | 0 to 14 pH |
|---|---|
| Operating Range 2714-HF | 0 to 12 pH |
| Operating Range 2715, 2717 | ± 2,000 mV |
| Pipe Size Range ½ in. and up | use Signet installation fittings from ½ to 4.0 in. (use pipe adapter in pipes over 4 in.) |
| Mounting In-line pipe mounting | > 5° to the horizontal plane, Submersion with 3-2720 preamplifier requires ¾ in. NPT or ISO 7-1/R 3/4 in. Male threaded extension. |
| Mounting Efficiency | > 97% @ 25 °C (77 °F) |
| Mounting pH Response Time | < 5 secs. for 95% of signal change |
| ORP response time | application dependent |
| Reference Electrolyte Solidified Acrylamide Gel | 3.5M KCI , 0.1 M KCl (2716-DI only) |
| Reference Secondary junction: | Nylon filament |
| Reference Element | Ag/AgCl |
| Primary Functions 2714/2715 | Flat surface resists fouling |
| Primary Functions 2716/2717 | Bulb surface for general use |
| Primary Functions 2714-HF | Extended use in applications with trace hydrofluoric acid (<2%) |
| Primary Functions 2716-DI | Extended use in pure waters (<100 μS) |
| Temperature Sensor | 3K Balco (3000 ohm = 25 °C) |
| Response Time, τ | 2714: 140 secs. , 2716: 196 secs. |
| Body | CPVC |
| O-rings | FPM |
| O-ring junction | Porous UHMW Polyethylene |
| Sensing Surface | glass membrane (pH), platinum (ORP) |
| Operating Temperature 2716/2717 | 0 °C to 85 °C (32 °F to 185 °F) |
| Operating Temperature 2714/2715 | 10 °C to 85 °C (50 °F to185 °F) |
| Operating Temperature 2714-HF | 0 °C to 50 °C (32F ° to 122 °F) |
| Storage Temperature | -10 °C (15 °F) to 30 °C (86 °F) |
| Temperature and Pressure Limit | 6.89 bar @ 0 °C (100 psi @ 32 °F to 149 °F) , 4.00 bar @ 85 °C (58 psi @ 150 °F to 185 °F) |