GF Signet 3-2630-2 Free Chlorine Electrode

GF Signet 3-2630-2 Free Chlorine Electrode

GF Signet 3-2630-2 2630 Electrode. Mounting: DryLoc connection, O-ring Material: FKM, Wetted Material: PVC, PTFE, FKM, Nylon, Silicone
SKU:3-2630-2
Model Number: 3-2630-X
Price: $1409.00

GF Signet 3-2630-2 Amperometric Chlorine Electrode. Free Chlorine electrode, 0.02 to 2 ppm (mg/l)


WHERE TO BUY?
Overview
Specification
Accessories
Overview
Features:
  • Embedded memory chip accessible via the 9950-X Chlorine Controller
  • Quick assembly with the patented DryLoc® connector
  • Integrated temperature element for automatic temperature compensation
  • Separate drive electronics (Signet 2650), for easy electrode replacement without running new cable
Applications:

Residual Chlorine Monitoring:

  • Water Distribution
  • Ground Water
  • Surface Water
  • HVAC Applications (cooling water)
  • Boiler Feed Water
  • Food and Beverage
  • Swimming Pools
  • Aquariums
  • Water Parks
Specification
TypeElectrode
Series2630
MountingDryLoc connection
MaterialsCPVC
Polarization Source2650 Amperometric Electronics
Free Chlorine Membrane MaterialPTFE
O-ring MaterialFKM
Working ElectrodeGold
Counter Reference ElectrodeSilver halide
Wetted MaterialPVC, PTFE, FKM, Nylon, Silicone
Repeatability&plusmn0.08 ppm (mg/l) or 3% of selected range whichever is less
Slope15 to 60 nA/ppm (mg/L) @ 25&degC
Response Time, T90< 2 minutes
Accuracy< &plusmn3% of electrode signal after calibration
Resolution&plusmn0.5% of electrode range
New, first start-up4 hours maximum before calibration
Subsequent Start-ups2 hours maximum
Temperature ElementPt1000
Free Chlorine Range0.02 to 2ppm (mg/l) 0.05 to 5ppm (mg/l) 0.1 to 20ppm (mg/l)
Free Chlorine pH Operating Range5.5 to 8.2 pH
Operating Temperature5&degC to 45&degC 41&degF to 113&degF
Membrane0.48 bar @ 25&degC (7 psi @ 77&degF)
Min Flow Velocity Across Membrane Surface15 cm/s (0.49 ft/s)
Max Flow Velocity Across Membrane Surface30 cm/s (0.98 ft/s)
SensitivityClO2 , ozone, bromine
Chemical Compatibility< 50% ethanol/water, < 50% glycerol/water
System Temperature-10&degC to 60&degC 14&degF to 140&degF
Storage Temperature-10&degC to 60&degC 14&degF to 140&degF
Relative Humidity0 to 95% indoor/outdoor non-condensing to rated ambient
StandardsCE, FCC
ApprovalsRoHS compliant, China RoHS
Compatible Flow Cells3-4630.392 (159 001 690) 3-3610-1 (159 001 683) 3-3610-2 (159 001 684)
Accessories
GF Signet 3-2630.398 2630 Chlorine Sensor Maintenance Kit
GF Signet 3-2630.391 2630 Free Chlorine Electrolyte
GF Signet 3-2630.394 Chlorine Sensor PTFE Membrane