GF Signet 3-2726-00 DryLoc Bulb pH Electrode
Bulb pH sensing tip with Pt1000 temperature element
SKU:3-2726-00
Model Number: 3-2726-XX
Price: $449.00
The 3-2726-00 uses a bulb-style pH sensing tip with a Pt1000 temperature element. It connects through DryLoc and is specified with a 3/4 in. NPT process thread.
GF Signet 3-2726-00 DryLoc Bulb pH Electrode
The 3-2726-00 uses a bulb-style pH sensing tip with a Pt1000 temperature element. It connects through DryLoc and is specified with a 3/4 in. NPT process thread.
This is the bulb pH version, whereas the 3-2724-00 has a flat pH tip and the 3-2725-60 measures ORP. Match the electrode geometry and compatible electronics to the measurement installation.
| Efficiency | >97% @ 25 °C (77 ° F) |
|---|---|
| Operating Range pH | 0 to 14 pH |
| Compatibility | 2750 Electronics, 2760 Preamplifier |
| pH Temperature Sensor | PT1000 versions are compatible with Signet 2750 pH/ORP Sensor electronics for connection to a PLC or to the Signet 8900 MultiParameter Controller. , 3 KΩ Balco versions are compatible with the Signet 2760 pH/ORP preamplifier for connection to the Signet 5700 pH/ORP Monitor and the Signet 8750 pH/ORP |
| Process Connection | ¾ in. NPT , ISO 7/1-R 3/4 , Mounts into Signet fittings |
| Wetted Materials | pH: Ryton® (PPS), glass, UHMW PE, FPM, ORP: Ryton® (PPS), glass, UHMW PE, FPM, Platinum |
| Operating Temperature Range | -10 °C to 85 °C (14 °F to 185 °F) |
| Operating Pressure Range | -10 °C to 65 °C (14 °F to 149 °F): 0 to 100 psi / 65 °C to 85 °C (149 °F to 185 °F), linearity derated 100 psi to 58 psi / *Best performance for 2726-HF sensors is above 10 °C (50 °F) |
| Recommended Storage Temperature | The best storage temperature for the 272X pH and ORP electrodes is 0 °C to 50 °C (32 °F to 122 °F) , The electrode glass will shatter if shipped or stored at temperature below 0 °C (32 °F) , The performance life of the electrode will shorten if stored at temperatures above 50 °C (122 °F) |
| In-line Mounting | Use the sensor threads , Use a Signet standard fitting up to 4 in. , Sensor can be mounted at any angle Submersible Mounting: , Use threads on models 2750 or 2760 , Requires ¾ inch NPT or ISO 7/1-R 3/4 male threaded liquid tight extension conduit. |
| Standards and Approvals | Manufactured under ISO 9001 for Quality and ISO 14001 for Environmental Management |